UV Lamp
Performance Over Time |
Authors: Stephen B. Siegel, Peter
Mandellos, David Luster
UV Process Supply, Inc. Chicago, Illinois
USA
| Summary |
|
In this study, we look
at the UV Lamp itself as the variable of primary concern in the UV curing process. We
have performed two tests: one which displays graphically how an individual
UV lamp
degrades overtime, and another which shows how one UV lamp compares to another. We will
discuss all these results in terms of intensity at critical wavelengths. |
| |
| Introduction |
|
UV ramp performance has
classically been characterized by the use of two distinct methods. Many production
facilities employ indirect mechanical testing of the resultant products, which makes
assumptions about the output of the UV Lamp based on cure characteristics of the
product, and which may have a multitude of undiscovered variables attached such as heat
effects, oxidation effects, or I.R. effects. The second method, a more direct (but
limited) radiometric measuring device, produces a "Snapshot" of the lamp's
performance at the moment of its use. Collecting data as it moves through a process,
radiometers sample conditions at a fixed rate, but often are only used at the beginning of
a production run, or maybe periodically to "Spot Check" the process on important
jobs. Yet field operations still fail to produce consistent results utilizing these
methods.
Our recent development of a Continuous Lamp Monitor,
the LM-9000, has enabled us to see why many of these failures occur, can provide insight
into what steps to take to eliminate or curtail the unwanted effects, and generate
consistent production output. |
| |
| Real World Concerns |
|
Many companies involved with
UV
Curing continue to experience problems with their curing systems, either intermittent
product failures, or entire runs which do not meet Quality Control standards even though
they were "produced under the same conditions" as jobs which passed those same
standards. The introduction of other variables into the curing equation is the obvious
answer, but isolating and identifying those variables is nearly impossible with the tools
at hand.
With only the mechanical tests of failed products,
or even the radiometer readings at job start to provide raw data, assumptions about cause
and effect are called on to fill in the holes in our knowledge. What is called for is a
means to Continuously Monitor the output of the lamp. |
| |
| The Test Instrument |
|
Continuous Monitoring of
UV Lamp
performance was accomplished through the use our LM-9000 Lamp Monitor, an analytical tool
for directly measuring the output of UV Lamps while on-line, operating inside the curing
system. It is a computer (Windows P.C.) hardware and software based Data Acquisition
System comprised of an optical probe, fiber optic cabling, an optical bench, sensor unit
and the Lamp Monitor software.
The Optical probe, which has a rated max.
temperature of 400~C, is mounted within the Lamp Reflector Housing or clipped directly
onto the lamp. It is connected via fiber optic to an optical bench, mounted on an internal
P.C. card or in an external case (attached via serial cable). This unit splits the
incoming light "signal" from the fiber optic into its various components
(Wavelengths).
An array of 1100 sensors then converts this
wavelength information into electrical signal voltages which the Lamp Monitor Software
measures and manages. This software package displays the Wavelength Intensity Distribution
between the 200nm and 450nm wavelengths in real-time, sampled approximately every 5
seconds. All wavelengths within the range (all 1100 data points) were examined at each
sampling.
Particular wavelengths can be selected for special
attention, such as to generate "Events" or to signal various simulated phases of
actual production runs (Start, Stop, Change rower Settings, etc.).
The Life of Lamp Test (single lamp) data was
collected from a 10" arc length lamp, at 750 watts per inch, over it's actual
life of 408 hours (18 days) continuous use before failure.
The Lamp Comparison Tests data was collected from 14
lamps, all 20" arc length, at 300 watts per inch, each lamp run for 15-20 minutes,
but sampled only after Lamp Stabilization had occurred. Two (2) were brand new, the others
were returned from Production Environments after 1000 hours of use. Some of these lamps
were of Foreign manufacture, while others were of Domestic origin. |
| |
| The Baseline of A Lamp |
|
Every UV lamp produces specific
wavelengths of light associated with the materials inside the lamp (i.e. mercury, metal
halides, et.al.). Each lamp produces its own Wavelength Intensity Distribution over its
given range (usually between 200nm and 450nm) which can be viewed as it's signature.
Certain frequencies within this range are important
to the curing process, and the others are either unnecessary, or harmful, to that specific
process. photo-Initiators can respond, to greater or lesser degree, at various frequencies
(See Table 1). Once power has been applied to a lamp, it goes through a
"Warm-Up" period, during which time it's output levels are unstable. This makes
the lamp an unreliable curing instrument until Stabilization occurs.
We observed that this Warm-Up period is different
for every lamp and power supply combination, lasting for between S and 15 minutes, during
which intensity levels of the various wavelengths fluctuate wildly. Some wavelengths can
reach 15000 of their operating maximum before settling down, others do not approach 10000
until lamp temperature reaches some critical point.
Once the lamp has reached this Stabilized state, we
save all associated values to form the "Lamp Baseline". By saving
the signature of the lamp when it is "New" or newly installed into the curing
system, we have a permanent set of Baseline values to use for comparison at any time in
the future. |
| |
| Energy Output Events |
|
From Lamp Start, when energy output
of all wavelengths was Zero, through the Warm Up period, to the time when output
characteristics appear to Stabilize, the particular wavelengths showing the highest
intensities (or peaks) changes as the lamp proceeds to its operating state. Radiometers
give readings showing Total Energy output (in mJ/cm2 ) on a narrow range of
wavelengths, presumably some of those same high-intensity peaks, sampled at a given rate,
such as 40 samples per second. Such devices may show the same reading even though the
actual wavelengths of light being output is rising or falling within the range.
As an example, from Lamp Start until Three minutes
had elapsed, one lamp displayed peaks at 350nm, which after Four minutes became two lower
peaks both above and below 350nm. Total energy of the individual 350nm peak was about the
same as the two lower adjacent peaks. This type of behavior would not be differentiated by
the "Snapshot" method of a radiometer: both would have given similar results
even though the actual output frequencies were different (see figure 1).
By using the Baseline of a lamp obtained previously,
and comparing it to the present state of that lamp (which was just started), a real-time
display of percentages at the six selected wavelengths allowed us to see when the lamp
Stabilized. This Stabilized state was defined by viewing this percentage display and
noting when the intensity values no longer increased.
Once the intensity levels of the various wavelengths
had Stabilized, the Warm-Up phase ended and the simulated Job Start would begin, as in a
production environment. We set up our instrument to save a Job Baseline of all
wavelengths, the simulated job would then proceed with subsequent samplings occurring at
regular (under 5 second) intervals.
This regularly sampled data could be saved at either
regular intervals or saved only upon an "Event", such as when any monitored
wavelength's intensity level fell below 2500 or 5000 of it's baseline value. |
| |
| Life of Lamp Test |
|
We monitored one lamp (750 watts
per inch) running continuously for 18 days (408 hours) until its failure, in order to
determine an individual lamp's performance characteristics (see figure 2). As a lamp
experiences continuous use, we expected to find that Lamp Intensity in general degrades
over time. As the contents within a lamp experience changes in pressure, as the electrodes
inside the lamp decays and deposits material on the interior quartz, and possibly as the
quartz itself changes with temperature, the lamp's overall output degrades.
What we observed, however is that certain
wavelengths exhibit straight line decline in intensity while others show oscillating
declines, with peaks, valleys, plateaus and dips having a generalized trend downward.
Those wavelengths with greatest intensity tend to exhibit this agitated decline while
wavelengths of lower intensity tend to exhibit more straight line behavior (see figure 2).
Our most startling discovery to date (see figure 3)
is that there does not appear to be any uniformity, or even any simple trend, to this
degradation. Different wavelengths degrade at different rates. The percentage of
energy loss over the life of the lamp shows clearly that 254nm suffered the steepest
decline in intensity while 365nm held up the best overtime.
We have observed that subtle variances in power
supplied to a lamp affects its output. When a factory up the street starts up or shuts
down, an "Event" would be triggered, alerting us that lamp behavior had changed
(see figure 2).
Similarly, changes in ventilation method, as well as
the individual operating temperature of the lamp, had profound effects on the intensity
distribution of that lamp. If ventilation air is blown across a lamp, the temperature of
that air may cause the intensity of specific wavelengths to drop significantly from the
lamp's Baseline. Natural convection was chosen for the tests we performed, to help
eliminate this variable from our results. |
| |
| Lamp Comparison Tests |
|
In order to perform a broader
examination of lamp characteristics, we monitored 14 lamps of the same specification:
20" arc length, 300 watts per inch. Some were new, some had over 1000 hours of
service, some were made in the United States, while others were of foreign manufacture.
All lamps were installed into the same testing
enclosure, provided with the same line current and power supply, and run under the same
conditions. The same testing instrument was used to obtain all the results.
The intensity levels of a range of similar lamps at
a particular wavelength varies. When looking at the six different wavelengths we were
paying special attention to (see figure 4), each lamp in the series displayed distinct
intensity levels at each wavelength, though they do tend to congregate around a certain
level. Output measured at 254nm tends to congregate around an intensity index value of 250
whereas output measured at 365nm tends toward an intensity index value of 3000.
The degree of separation lamp-to-lamp, between index
values of individual lamp intensities at these pre-selected wavelengths, can also be seen
to vary widely. Some of the selected wavelengths held tightly together, with less than 500
variation lamp-to-lamp, while others changed considerably.
Further, some "used" lamps (with over 1000
hours in service) produced higher intensity levels at critical frequencies than did
comparable "brand new" lamps, or used lamps with higher or lower hours (again,
see figure 4). We may be observing variances in internal lamp conditions, where a
particular lamp's electrodes have decayed or changed conductance.
We had seen that with any given lamp, the individual
wavelengths degrade at individual rates, some displaying marked fall off (4000 or more
from Baseline) while other wavelengths remain strong (within 1500 - 2000 of Baseline).
Some uniformity in degradation of lamp performance between similar lamps would be
expected, given general entropy effects and the observed variances in specific wavelength
output.
However, we have seen no true pattern of decay
between similar lamps manufactured by the same company, much less by similar lamps
produced by different lamp manufacturers. Lamps produced on the same day, to the same
specifications, by the same manufacturer exhibit different operating characteristics
overtime, even within the same curing system.
At different power supply settings on the same lamp,
we have also observed non-linear variances in intensity within the selected wavelengths
(see Figures 5a through 5d and 6a through 6d) the same group of lamps were tested at High
and Medium power settings, 300 watts per inch and 200 watts per inch respectively. The
notion that decreasing the power level decreases the intensity of all the individual
wavelengths within a lamp proportionately, would produce a set of graphs that look
identical, except for the intensity level (i.e.: the height of each bar in the bar graph,
or the starting point of the line graph).
While the graphs are similar, they are not quite
proportional, with some lamps exhibiting a noticeable difference in output wavelength
intensity when compared to others which did perform more proportionately. Again, new lamps
did not distinguish themselves from used lamps in these regards; they were not any more
linear in behavior than the lamps with 1000 hours. |
| |
| Conclusions |
|
Many factors influence the level of
cure within a process. Higher power lamps, newer coating formulations, new trends in
cooling and I.R.~UV filtering have all contributed greatly to a more stable, and hence,
reproducible finished product. The improvement in control systems and tools such as
radiometers, helps promote ever higher Quality Control standards. But, if the underlying
mechanism for obtaining cure, the UV lamp itself is not providing long-term, stable
output at the required wavelengths, the entire process can be compromised, or even
inhibited.
Finished product failure could result, even though
all the periodic spot-checks indicated a successful production run. Periodic spot-checks
only display state at the moment they are performed, while the curing process requires
on-going, continuous monitoring of lamp output to insure thorough cure of the finished
product.
 | We have observed the individual lamp performance
over time and found various wavelengths loosing intensity at different rates over the life
of the lamp.
|
 | We have observed that similar lamps do not have
the same Wavelength Intensity Distribution signatures.
|
 | We have observed that a lamp's Baseline
performance characteristics and subsequent degradation from Baseline are not linear for
every frequency within it's output range.
|
 | We have observed that a lamp's Baseline
performance characteristics are not linear for every frequency within it's output range at
different power settings.
|
 | We have observed that lamps with the same arc
length, same power rating and even same manufacturer produce different Wavelength
Intensity Distribution signatures.
|
 | We have observed significant changes from Baseline
during the course of a job, or over the course of several weeks of operation, and in
general over the life of the individual lamp.
|
All of these incidents lead to an unstable-stable,
not very reproducible end result. Thus the characteristic Wavelength Intensity
Distribution signature of each lamp plays a major role in it's long term use as a reliable
curing instrument.
Fluctuations in power, both the external (utility
supplied) line power characteristics as well as the individual lamp power supply, may
certainly be responsible for some of the observed effects. Similarly, the amount of
ventilation, and how it is supplied to the individual lamps may have observable effects on
the operating behavior of the lamps. Further testing in these areas will be done over the
next few months to determine their relative impact on the results obtained so far.
Continuous Lamp Monitoring, and event driven
responses to run-time conditions, is clearly called for, in order to reduce the occurrence
of problems. Only through real-time examination of lamp output can we avoid the pitfalls
of current production methods, because periodic spot-checks are rarely performed at the
moment they are needed, when the lamp is not producing the required intensity at the
required wavelengths. |
|
| |
|